LBL04002: Difference between revisions
mNo edit summary |
|||
Line 36: | Line 36: | ||
RO water needs to be regularly topped up, into the hatch on the top of the machine. | RO water needs to be regularly topped up, into the hatch on the top of the machine. | ||
The autoclave tape should be used to seal items to be sterilized and is stored on the shelf nearby. | The autoclave tape should be used to seal items to be sterilized and is stored on the shelf nearby. Order replacement from [https://www.sigmaaldrich.com/catalog/product/aldrich/br61750?lang=en®ion=GB Sigma] or appropriate similar product. | ||
Revision as of 11:48, 27 November 2014
Using the Autoclave for Media Preparation
Author: Tolland | Approved by: S. Thompson | SOP No. LBL04002 |
Signed: | Signed: | Effective from: |
Date: | Date: | Last edited: |
1. Purpose
This describes the operation of the Melag EuroKlav 29V-S. The autoclave is used for two main procedures:
i) sterilization of media, glassware and autoclavable equipment prior to another protocol ii) Deactivation of GMOs, and GMO exposed waste and equipment.
2. Scope
This should be observed every time the autoclave is used within the LBL BioLab.
3. Responsibilities
The operator of the autoclave is responsible for their own safety and that of others in the vicinity during its use. The particular risks presented in using this piece of equipment are primarily due to the high temperatures generated, steam and high pressures.
The operator is responsible for ensuring complete sterilization of containers, glassware, tubes, media and equipment. To ensure environmental microbes do not contaminate the target microbe strain.
4. Materials
The main consumables for the autoclave are RO water and autoclave tape.
RO water needs to be regularly topped up, into the hatch on the top of the machine.
The autoclave tape should be used to seal items to be sterilized and is stored on the shelf nearby. Order replacement from Sigma or appropriate similar product.
5. Related documents
http://www.phytotechlab.com/pdf/SterilizingNutrientMedia.pdf
6. Definitions
None given
7. Procedures
7.1 Sterilization of liquid media
7.1.1 Use the * Sterilize the load using the Quick-Programme (15 psi/121 C/15 mins)
7.2 Deactivation of liquid and solid GMO waste ==
7.2.1 Transport the waste to the autoclave in a non-drip autoclavable tray 7.2.2 Ensure there are no folds or containers which can trap air 7.2.3 Sterilize the load using the Universal-Programme (2.1 bar/134 C/15 mins) 7.2.4 once the programme has run, the waste should be transferred to the chemical kill bin, for final deactivation
8. Resources
- Operating manual: http://www.eickemeyer.nl/images/products/2281/500032_Melag_Euroklav_23VS.pdf