'''1. Purpose'''
'''1. Purpose'''
This describes the operation of the Melag EuroKlav 29V-S. The autoclave is used for two main procedures:<br>
This describes the operation of the Melag EuroKlav 29V-S. The autoclave is used for two main procedures with respect to GMO deactivation:<br>
     i) sterilization of media, glassware and autoclavable equipment prior to another protocol
     i) sterilization of glassware and equipment for reuse
   ii) Deactivation of GMOs, and GMO exposed waste and equipment.
   ii) Deactivation of small volumes of GMO exposed media, and GMO exposed waste and equipment.
  iii)


'''2. Scope'''
'''2. Scope'''
Line 51: Line 52:
7.1 Daily Test procedure
7.1 Daily Test procedure


Before using the autoclave for GMO deactivation, the operator should run a test strip through the autoclave to check that it is working to specification.
Before using the autoclave for GMO deactivation each day, the operator should run a test strip through the autoclave to check that it is working to specification.


The test strip should be stapled into the lab book, with a note indicating the operator and the date/time the test was run, along with any observations.
The test strip should be stapled into the lab book, with a note indicating the operator and the date/time the test was run, along with any observations.


7.1.1 Sterilize the load using the Universal Programme (28 psi/134 C/30 mins)
7.1.1 Unwrapped equipment and glassware with no internal cavities can be sterilized using the Quick Programme (28 psi/134 C/3 mins)
 
[[File:Screenshot-quick-programme.png]]
 
7.1.2 Materials such as a solid and liquid media can be sterilized using the Universal Programme (28 psi/134 C/30 mins)
 
[[File:Screenshot-universal-programme.png]]
 
7.1.3 Large volumes of liquid or solid media should be processed with extra care.
 
The autoclave should be preheated;
 
[[File:Screenshot-preheat.png]]
 
If possible, the media should be split into several autoclaves sessions. The load should be run under the Universal programme
 
[[File:Screenshot-universal-programme.png]]