| Currently working.
| |
|
| |
|
| http://www.frankshospitalworkshop.com/equipment/documents/centrifuges/service_manuals/Jouan_CR3i_-_Service_manual.pdf
| | The Jouan C3i is a laboratory centrifuge located in the first cupboard in the biolab. |
| | |
| | It has capacity for 6 x 50mL falcon tubes, and has a maximum speed of 10,000 rpm. |
|
| |
|
| The tachometer works (at low speed), the microswitches on the lid interlock switch on and off and the control electronics and front panel seem to work. | | The lid has an interlock requiring it to be powered on to access the rotor cups. |
|
| |
|
| There is a connection diagram for the main board on page 2-11 (page 20 in the pdf linked above). The motor drive is powered directly from the mains, separately to the rest of the main board.
| | This equipment requires an induction if you do not have normal experience with these types of centrifuge. |
|
| |
|
| Connector J12 (supplying mains power to the motor drive) is badly damaged and I have removed it. This is a three way Panduit brand IDT connector. I think it is an ITW Pancon brand MAS-CON 3.96mm connector (eg. http://uk.farnell.com/itw-pancon/ce156f22-3-d/wire-board-conn-female-3pos-3-96mm/dp/1418234).
| | http://www.frankshospitalworkshop.com/equipment/documents/centrifuges/service_manuals/Jouan_CR3i_-_Service_manual.pdf |