|maintainers=Biohackers
|maintainers=Biohackers
}}
}}
Currently broken. We'd like to fix it.
Currently working.


[[user:mentar|Mentar]] had a look at it. What we know:
* It switches on but won't spin
* Motor is ok
* Problem is with the electronics
* Could be to do with the locking mechanism (machine thinks it's not locked when it is and hence won't start)
* It's definitely not that, it went BANG! and blew the fuse
[[user:amc|Andy]] had a look at it (2012/07/27):
This appears to be the service manual for it:
http://www.frankshospitalworkshop.com/equipment/documents/centrifuges/service_manuals/Jouan_CR3i_-_Service_manual.pdf
http://www.frankshospitalworkshop.com/equipment/documents/centrifuges/service_manuals/Jouan_CR3i_-_Service_manual.pdf


Line 31: Line 21:


Connector J12 (supplying mains power to the motor drive) is badly damaged and I have removed it. This is a three way Panduit brand IDT connector. I think it is an ITW Pancon brand MAS-CON 3.96mm connector (eg. http://uk.farnell.com/itw-pancon/ce156f22-3-d/wire-board-conn-female-3pos-3-96mm/dp/1418234).
Connector J12 (supplying mains power to the motor drive) is badly damaged and I have removed it. This is a three way Panduit brand IDT connector. I think it is an ITW Pancon brand MAS-CON 3.96mm connector (eg. http://uk.farnell.com/itw-pancon/ce156f22-3-d/wire-board-conn-female-3pos-3-96mm/dp/1418234).
The symptoms I've seen are consistent with the motor drive not getting any power (due to the damaged connector) or being otherwise broken (and hence damaging the connector).
I suggest:
* Step 1: replace J12, either with the same type (if we can find a source) or with another 3.96mm connector of suitable specification.
* Step 2: carefully test the motor drive circuitry.
* Step 3: if the motor drive circuitry appears to work test the whole centrifuge.